E6000 Quick Reference Guide

WARNING! - Before proceeding with the combustion analysis remember to set the proper fuel. 7/08/14 10:00 7/08/14 10:00 ...

0 downloads 149 Views 876KB Size
WARNING! - Before proceeding with the combustion analysis remember to set the proper fuel. 7/08/14 10:00

7/08/14 10:00

Configuration Analysis

Configuration

Compliant with: EN 50379-1, EN 50379-2 Features

7/08/14 10:00

• Combustion, Flue gas & Emissions analysis • Calculating of stack heat loss and efficiency • CO environment measurement • Measuring differential pressure • Draft measurement • Pressure measurement in the gas supply pipe • Store Smoke values, calculating mean value

Configuration Fuel Natural gas

Analysis

Instrument

Fuel

OK

Condensation

Operator

Alarms

O2 reference

NOx/NO ratio

Information

Diagnostic

Measure units

Autozero

Propane

OK

COMPLETE MANUAL ON USB FLASH DRIVE

0 E6 00

L.P.G. Butane Diesel oil Fuel oil

The magnets in the back of the instrument can damage credit cards, hard drives, mechanical watches, pacemakers, defibrillators and other devices proven sensitive to magnetic fields. It is recommended to keep the instrument at a distance of at least 10” away from any of these devices.

Propane-Air Language

Measures list

Restore

7/08/14 10:00

7/08/14 10:00

Combustion analysis

O2

4.2

Mode

CO2

9.3

Memory

1.25

Samples

10

Interval

60

%

Ex. Air T flue

190.1

°F

T air °F

ΔT °F

Loss tot

7/08/14 10:00

Memory Save

Combustion analysis

%

Biogas

Air temp.

data logger 1

s

Eff. tot

CO2

9.3

%

OK

Ex. Air

1.25

T flue

190.1

°F

15.4

T air

74.7

ΔT

15.4

°F

74.7

°F

Loss tot Eff. tot

o

1 120

7/08/14 10:00

Memory Average analysis

Print Report

4.2

Memory

1

CO2

9.3

Analysis

average

Ex. Air T flue °F

T air °F

ΔT °F

Loss tot %

Eff. tot %

1.25

Model

partial

Memory

OK

Analysis WARNING In average Model

190.1 15.4

1

Printing. Please wait...

F1: stop

74.7 8.6

partial

Date: 7/08/14 Time: 10.10 Fuel: Natural gas Altitude: 0 m R.H. air: 50 % O2 4.2 ٪ CO2 9.3 ٪ Ex. Air 20 ٪ T flue 374.2 °F T air 60.4 °F ∆T 313.8 °F Loss tot 8.6 ٪ Eff. Tot 91.4 ٪ CO 148 ppm NO 40 ppm NOX/NO: 1.03 NOX 41 ppm CO amb 0 ppm Draft: 0.05 inH2O Smoke: 3 1 2 Aver n. : 2

91.4 OK

• To view different test measurements, use '

ZERO the pressure sensor Updates the measure Saves the current measurement or analysis in the memory location selected from the ‘Select Memory’ menu

Enters the Print menu

Starts the report ticket Enters the Print menu

o

Quits the current screen

Stops the flue gas analysis Enters the modify mode for the selected parameter

Enters the measurements menu

7/08/14 10:00

Print Report

O2

O

Enters the Memory menu

Turns the instrument on/off

7/08/14 10:00

%

Activate the context keys shown on the display

Previes the combustion analysis

In case the automatic print mode is set then the average analysis printing start automatically. Conversely, after the sampling of the third analysis, the average analysis is shown, so that it can be sent to the printer as explained in the following.

%

CONTEXT KEY FUNCTION

FUNCTION

Enters the Configuration menu

91.4

%

OK

Automatically saves the first sample when the set interval expires, and so on until the third sample.

CONTEXT KEYS

KEYS

8.6

%

91.4

%

4.2

%

8.6

%

O2

KEYBOARD FUNCTIONS

OK

Confirms the settings

Esc

Cancels the change/operation returns to the previous screen

and

Zooms the screen Selects and/or modifies

Shows the details for the selected parameter

Confirms data

‘Find’ function; starts a quick search for the analysis to recall

E Instruments respects the nature and the environment, therefore provides this quick user guide to avoid any unwanted waste of paper. However the complete manual for use and maintenance of the instrument is already available in the USB stick (included),. Respect your environment: think before printing the full manual on paper.

F1

' key to scroll through the various test measurement screens. Xair = Excess Air Tg = Flue Gas Stack Temperature Ta = Ambient or Incoming Air Temperature ΔT = Differential Temperature (NET Temperature) Eff. tot = Total Combustion Efficiency Loss tot = Stack Losses

402 Middletown Blvd, Suite 216, Langhorne, PA 19047 USA Tel: (215) 750-1212; Fax: (215) 750-1399 [email protected] - www.E-Inst.com K600000000EJ 028006 020217

7/08/14 10:00

7/08/14 10:00

It is mandatory to perform the instrument autozero in fresh, clean outdoor air.

Measurements Draft

Memory

P diff

AA AL05 Save

Average

0.001

Draft

Select

Data logger

Delete

Usage %

inH2O

Before zeroing the pressure it is mandatory to REMOVE the probe from the chimney.

7/08/14 10:00

Measurements Ambient CO

CO

0

Ambient CO

O

ppm

CO max

AAC TA03A O

ZERO Draft Sensor

0 ppm

7/08/14 10:00 Select

Data logger

OK

OK

Measurements Draft

P diff

AAC KP01 AA SA08

AA SF--A

WARNING! • Be sure all connections are tight to assure accurate sampling. • It is important that water trap & filter be installed vertically to prevent damage from moisture & particles to sensors. • When conducting measurements, the water trap/filter assembly MUST be in a VERTICAL position. • When testing is completed, always drain the water trap with any condensation (after EVERY test!).

0.000 7/08/14 10:00

Memory Select Memory

E Instruments 402 Middletown Ste 216

Langhorne, PA 19047 Phone Boiler

During autozero DO NOT insert the gas probe in the chimney.

Combustion analysis UNI 10389 O2

4.2

CO2

9.3

% %

Ex. Air T flue °F

KEEP PRESSED FOR 2 SECONDS UNTIL IT BEEPS

T air °F

ΔT °F

Loss tot %

Eff. tot %

1.25 190.1

Date

inH2O

Memory Data logger 1

Customer Address

7/08/14 10:00

7/08/14 10:00

Mode

10

Interval

60

Print

auto

Memory

auto

O Save the measurement

View & Save the DRAFT measurement.

215-750-1212 xxxx 7/08/14

In "Select" can be i nser ted al l the customer data.

Connect the gas probe to the instrument and perform the measurement. Wait 5 minutes.

Data logger

Samples ppm

Once the pressure zeroing is completed insert the probe in the chimney and measure the draft.

7/08/14 10:00

In "Data logger", the user can define the anal ysi s, m em or y selection and printing modes.

7/08/14 10:00

Measurements Draft

P diff

Measurements Ambient CO

CO

- 0.020 inH2O

0 ppm

CO max

15.4 74.7

0

8.6 91.4

ppm

O K600000000EJ 028006 020217